Opened Outage Report
losing connection to receiver, problem persisted after replacing receiver front panel board
On Mar 30, installed Tripplite Isolator on site to help condition the incoming power. On Apr 7, moved all equipment to the office for bench testing. Still cannot connect to receiver, even with direct USB connection to AWG module inside chassis. Will try replacing the AWG module.
Status switched to 'In Progress'
The AWG module and front panel board were both ok when installed in a 25 MHz chassis. Something else inside the receiver is not working. Shipped the unit back to CODAR RMA #2253
An IOOS loan is not available at this time.
Installed rx 2005127 and tx 2005127 taken from CEDR
Status switched to 'Closed'